Command:
BP_350<CR><LF> turn BEEP signal for 350 ms
Response:
BP_OK<CR><LF> BEEP on
Caution! BEEP signal activated by BP command gets aborted if it is activated
by means of other devices: keyboard, touch panel, proximity sensors.
OMI - list available working modes
The command lists available working modes for a given device
Format: OMI <CR><LF>
Accessible responses:
OMI <CR><LF>
n_”Mode name
<CR><LF>
:
n_”Mode name
<CR><LF>
OK <CR><LF>
command completed, available working modes are
given
OMI_I <CR><LF>
command understood but not accessible at this
moment
Mode name parameter, working mode name, given in between inverted
commas. Form of the name is the same like on the balance display, it is given
in a currently operating language.
n parameter, decimal value specifying working mode number.
n 1 Weighing
2 Parts counting
3 Percent setup
4 Dosing
5 Formulation
6 Animal weighing
7 – Density
8 Density of solids
9 Density of liquids
10 Peak hold
11 Totalizing
12 Checkweighing
13 Statistics
14 Pippette calibration
Caution: working modes numbering is assigned to working modes names and
it is the same for each balance types.
Some balances in response to OMI may give working modes numbers without
names.
Example1
Command:
OMI <CR><LF> give available working modes
Seitenansicht 220