28. MASS CONTROL
(not available in standard balance version)
Working mode < Mass control> enables controlling mass of a product that is
automatically batched (dosed) onto the balance’s weighing pan by an automatic
feeder PA-02/H that is connected to the balance. The working mode
automatically controls a complete batch of tested product after determining its
quantity.
28.1. Working mode activating procedure
Procedure:
• select < Mass control> the software automatically returns to the main
window and displays pictogram in the upper bar,
• Automatically a single on-screen button appears in the bottom bar of the
display:
Start mass control process
28.2. Mass control global settings
Entering the mode settings is carried out by pressing the grey coloured
workspace and selecting an group of parameters named "Settings" < >,
which contains the following:
Request batch
number
The function orders requesting a batch
number before starting mass control
Batch quantity
Determines quantity of a batch, i.e. number
of product units to be controlled.
Low limit
A value exceeding which causes accepting a
measurement record by a balance.
High limit
A value exceeding which causes omitting a
measurement in the summary, e.g. if two
parts (pills) are simultaneously dispensed on
Print report
A parameter enabling for switching on/off
automatic printout of a report on completing
Printout
A parameter enabling selecting whether on
completing the control process the balance
should trigger a standard or a non-standard
28.3. Mass control process
Before starting the mass control process carry out the following steps: