MASS FRAME „P1” CR LF
MASS FRAME „P2” CR LF
- immediate response of mass value in basic
measuring unit from all weighing platforms
Frame format with mass from each weighing platform:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
P n space
stability
marker
space character mass space unit CR LF
Where:
n - weighing platform number
mass - 9 characters with right justification
unit - 3 characters with left justification
Example:
In case there are two weighing platforms.
S I A CR LFcommand sent from a computer
P 1 _ ? _ _ _ _ _ _ 1 1 8 . 5 _ g _ _ CR LF
P 2 _ _ _ _ _ _ _ _ _ 3 6 . 2 _ k g _ CR LF - command carried out,
immediate response is mass value from both weighing platforms in basic
measuring unit specific for each of the weighing platforms.
Send stable measurement result in current weighing unit
Format: SU CR LF
Accessible responses:
SU_A CR LF
SU_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable measurement
result
SU_I CR LF
- command understood but not accessible at this moment
SU_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in current measuring unit
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S U space
stability
marker
space character mass space unit CR LF
Example:
S U CR LFcommand from a computer
S U _ A CR LFcommand understood and in progress
S U _ _ _ - _ _ 1 7 2 . 1 3 5 _ N _ _ CR LF - command carried out,
response is mass value in current measuring unit.
Immediately send measurement result in current measuring unit
Format: SUI CR LF
Accessible responses:
SUI_I CR LF
- command understood but not accessible at this
moment