Accessible responses:
- command understood, new PROFILE activated
- command understood, an error in PROFILE name,
log in failed
- command not recognized (error in format)
The result release
Format: SS CR LF
Accessible responses:
SS_OK CR LF - command understood, realization in progress
The command simulates pressing PRINT button on an overlay, according to
settings selected for result release.
Internal adjustment/calibration
Format: IC CR LF
Accessible responses:
- command understood, realization in progress
- calibration completed
- command understood, realization in progress
- time limit of result stabilization exceeded
- command understood but not accessible at this
moment
Lock automatic internal adjustment
Format: IC1 CR LF
Accessible responses:
- command understood but not accessible at this
moment
- unavailable operation, e.g. verified balance
For verified balances the automatic internal adjustment cannot be locked.
For non-verified balances the command locks internal adjustment, it may be
unlocked either by IC0 command or by switching the balance off.
The command does not influence settings responsible for adjustment process
beginning.
Unlock automatic internal adjustment
Format: IC0 CR LF
Accessible responses:
- command understood but not accessible at this
moment
Kommentare zu diesen Handbüchern