- 169 -
Send stable measurement result in basic measuring unit
Format: S CR LF
Accessible responses:
S_A CR LF
S_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable
measurement result
S_I CR LF
- command understood but not accessible at this
moment
S_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
Frame format:
1
2-3
4
5
6
7-15
16
17
18
19
20
21
S space
stability
marker
space character mass space unit CR LF
Example:
S CR LFcommand sent from a computer
S _ A CR LF - command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF - command carried out, response
is mass value in basic measuring unit.
Immediately send measurement result in basic measuring unit
Format: SI CR LF
Accessible responses:
SI_I CR LF
- command understood but not accessible at this
moment
MASS FRAME
- response is immediate with mass value in basic
weighing unit
Frame format:
1
2
3
4
5
6
7-15
16
17
18
19
20
21
S I space
stability
marker
space character mass space unit CR LF
Example:
S I CR LFcommand sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ k g _ CR LF - command carried out,
immediate response of mass value in basic weighing unit
Immediately send measurement results from all weighing platforms
basic measuring units
Format: SIA CR LF
Accessible responses:
SIA_I CR LF
- command understood but not accessible at
this moment
MASS FRAME „P1” CR LF
- immediate response of mass value in basic