Possible answers:
SUI_I CR LF
- command comprehended but cannot be executed
SUI_A CR LF
MASS FRAME
- command accepted and in progress
- mass value in current unit is returned immediately
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S U I stability space sign mass space unit CR LF
Example:
S U I CR LF – computer command
S U I ? _ - _ _ _ 5 8 . 2 3 7 _ k g _ CR LF - command executed and mass
returned
28.4.10. Switch on continuous transmission in basic unit
Syntax: C1 CR LF
Possible answers:
C1_I CR LF
- command comprehended but cannot be executed
C1_A CR LF
MASS FRAME
- command comprehended and in progress
- mass value in basic unit is returned
Frame format:
1 2 3 4 5 6 7-15 16 17 18 19 20 21
S I space stability space sign mass space unit CR LF
28.4.11. Switch off continuous transmission in basic unit
Syntax: C0 CR LF
Possible answers:
C0_I CR LF
- command comprehended but cannot be executed
C0_A CR LF
- command comprehended and executed