Enter P1 CAL menu, next start P1-05
tSt_Co procedure. Before the test
activation set its parameters in
accordance with diagram and description,
to be found below.
Use navigating arrows to set specified test
temperature, press ENTER to confirm.
Next, enter tolerable (accepted by the
user) temperature errors.
Use navigating arrows to set specified
error value, press ENTER to confirm.
Next enter serial number of the adjustment
set, which will be used for the test.
Use navigating arrows to set the number,
press ENTER to confirm.
Upon confirmation of the number, program
will automatically start drying process until
the specified temperature is reached.
Time and sensor temperature information
will be displayed.
Temperature will remain unchanged for 8
minutes (like for temperature adjustment
process).